Care After Use: VarioSurg

Care After Use: VarioSurg

1. Auto Cleaning

2. Remove the handpiece cord and the irrigation tube from the unit.

After separating the irrigation tube from the VarioSurg handpiece cord, place the cord inside the sterilization cassette with the handpiece still attached. We recommend to leave the tube holders fastened to the cable.
After using the irrigation tube, dispose of it as medical waste.

3. Tips, Tip Wrenches and Tips Holders

4. Sterilization

Steam autoclave is recommended. Perform autoclave sterilization at temperatures up to 135oC. Example: 20 minutes at 121oC or 15 minutes at 132oC. Note that you cannot autoclave the control unit and the foot control.